Spin Echo-Based Families
Learn about the Spin Echo-Based Families, which are sequences using 180° RF refocusing pulses for robust signal formation.
Spin Echo (SE)
- Design: 90° excitation followed by a 180° refocusing pulse.
- Key characteristics: High robustness to B₀ inhomogeneity; flexible contrast via TE and TR.
- Clinical applications: General T1-, T2-, and proton-density imaging.
- Vendors: All vendors SE.
Turbo Spin Echo (TSE / FSE)
- Design: Multiple refocusing pulses following a single excitation.
- Key characteristics: Faster than SE with preserved contrast; reduced susceptibility artifacts.
- Clinical applications: Routine neuro and body imaging.
- Vendors:
- GE: FSE
- Siemens: TSE
- Philips: TSE
- Canon: FASE
- Fujifilm: FSE
Single-Shot Fast Spin Echo (SS-FSE)
- Design: Entire k-space acquired after one excitation using long echo trains.
- Key characteristics: Strong T2 weighting; highly motion-robust.
- Clinical applications: Pediatric imaging, fetal imaging, non-breath-hold abdominal imaging, MRA.
- Vendors:
- GE: SSFSE
- Siemens: HASTE
- Philips: SSh-TSE
- Canon: FASE
- Fujifilm: SSFSE
Inversion Recovery (IR) Spin-Echo
- Design: Inversion pulse followed by SE acquisition.
- Key characteristics: Strong T1 contrast control via inversion time selection.
- Clinical applications: Soft tissues, tumors, musculoskeletal imaging.
- Vendors: All vendors IR-SE.
Inversion Recovery Turbo Spin Echo (IR-TSE / STIR)
- Design: IR-prepared TSE with tissue-specific signal nulling.
- Key characteristics: Robust fat suppression independent of B₀; increased SAR.
- Clinical applications: STIR-based musculoskeletal imaging.
- Vendors: All vendors STIR.
Fluid-attenuated inversion recovery (FLAIR)
- Design: IR-prepared TSE with CSF-nulling inversion time.
- Key characteristics: Suppresses fluid signal while preserving T2 lesion contrast.
- Clinical applications: Brain imaging, white-matter pathology.
- Vendors: All vendors FLAIR.
3D Double Inversion Recovery Turbo Spin Echo (3D DIR-TSE)
- Design: Two inversion pulses followed by 3D TSE readout.
- Key characteristics: Dual tissue suppression; high lesion conspicuity.
- Clinical applications: Brain imaging, multiple sclerosis.
- Vendors: All vendors DIR-TSE.
Three Dimensional (3D) Turbo Spin Echo (3D TSE)
- Design: Slab-selective 3D TSE acquisition.
- Key characteristics: Isotropic resolution with high anatomical detail.
- Clinical applications: Brain, spine, joints, MRCP.
- Vendors:
- GE: CUBE
- Siemens: SPACE
- Philips: VISTA
- Canon: 3D FASE
- Fujifilm: 3D FSE