ARTICLE - How to set up Saturation Bands

You are now able to use Saturation Bands for placement with independent angles.

  • After you select the sequence that you want to use a sat. band on, open the Advanced tab.
  • Enable saturation bands to Yes.
  • Click on the Geometry tab.
  • Select No. sat. bands that you want to use.
  • Select which Selected sat. band you want to move
  • HOLD DOWN the letter s • The sat bands can be rotated and moved by clicking the white box in the middle of the sat. band
  • Sat. bands thickness(mm) for size adjustment of the bands. Both sat. bands will be the same thickness.
  • Work on the Selected sat. band that you have selected. When you are ready to move to the next group, make sure you change the Selected sat. bands that you will be working on.

*Tip: This is what it will look like when setting up a sat band on a Lumbar Spine Axial T2 Multi Slice scan

  • At this point you can Save the scan and Run the pulse.

Expected results:
                       Sat Band on Sagittal                                               Sat Band on Axial